PHOENIX
Received Mar 6, 2008 · Event occurred Jan 4, 2008
Report 1025943 · MDR key 1025943
Device
Generic name
Dialysis Machine
Manufacturer
Gambro Renal Products, IncModel number
PHOENIXCatalog number
*
Lot number
*
Product problems
- Device Reprocessing Problem
- Residue After Decontamination
- Device Displays Incorrect Message
Patient
11 YR
- Therapy/non-surgical treatment/delayed/prolonged
Narrative
Description of Event or Problem
DIALYSIS TREATMENT DELAYED FOR ALARM CONCENTRATE TYPE ERROR A PUMP 40. FOUND MINERAL BUILDUP ON CONCENTRATE LINE CONNECTORS. OPERATING WITHIN MANUFACTURER'S GUIDELINES AFTER VINEGAR RINSE.