inforMED
MalfunctionFPS

FEEDING TUBE

Received Jul 7, 2020 · Event occurred Jun 20, 2020

Report 2245270-2020-00054 · MDR key 10239593

Device

Generic name

Feeding Tube

Manufacturer

Vygon

Model number

361.062

Lot number

UNKNOWN

Product problems

  • Material Too Rigid or Stiff
  • Material Deformation
  • Material Too Rigid or Stiff
  • Material Deformation

Patient

0 YR

  • No Known Impact Or Consequence To Patient
  • No Information
  • No Clinical Signs, Symptoms or Conditions
  • No Known Impact Or Consequence To Patient
  • No Information
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THIS IS THE SECOND OCCURENCE OF THIS COMPLAINT. FOR ADDITIONAL INFORMATION PLEASE REFER TO 2245270-2020-000049. THE FAILED SAMPLE WAS RETURNED TO VYGON AND WILL BE EVALUATED AS PART OF THE COMPLAINT INVESTIGATION. THE RESULTS OF THIS INVESTIGATION ARE STILL PENDING, AND WILL BE COMMUNICATED TO FDA WITHIN 30 DAYS OF COMPLETION.

Description of Event or Problem

6 FR PVC TUBES HARDENING DURING PATIENT DWELL.

Additional Manufacturer Narrative

THERE WERE 2 OCCURRENCES OF THIS ISSUE, THE DETAILS FOR THE EVENTS CAN BE FOUND IN THE FOLLOWING REPORTS: 2245270-2020-00049, AND 2245270-2020-00054. THE COMPLAINT WAS FORWARDED TO OUR PARENT COMPANY IN FRANCE FOR THEIR EVALUATION. THE INVESTIGATION SUMMARY IS AS FOLLOWS: ANALYSIS OF THE SAMPLES REVEALED HARDENING OF THE DISTAL END. THIS IS A WELL-KNOWN PHENOMENON OF PVC TUBES WHICH MAY HAPPEN WITH GASTRIC FLUIDS, DEPENDING ON PATIENTS. NEVERTHELESS, IT HAPPENS VERY RARELY. FOR THIS PATIENT, WE RECOMMEND SWITCHING TO -PUR- POLYURETHANE FEEDING TUBE. SINCE NO BATCH NUMBER WAS PROVIDED, IT WAS NOT POSSIBLE TO REVIEW THE BATCH HISTORY RECORDS. THE COMPOSITION OF THE PRODUCT HAS NOT CHANGED. WE ALSO DO NOT HAVE A RECORD OF SIMILAR CLAIMS IN THE PAST 3 YEARS. CORRECTIVE ACTION. VYGON FRANCE RECOMMENDS SWITCHING TO -PUR- POLYURETHANE FEEDING TUBE. NO FURTHER CORRECTIVE /PREVENTIVE ACTION IS INITIATED AT THIS TIME. HOWEVER, BOTH VYGON USA AND FRANCE WILL CONTINUE TO MONITOR THIS ISSUE.

Description of Event or Problem

6 FR PVC TUBES HARDENING DURING PATIENT DWELL.