MONITROL 15

Received Mar 21, 2008 · Event occurred Mar 5, 2008

Report 2126677-2008-00023 · MDR key 1019857

Device

Model number

45296001

Catalog number

NA

Lot number

NA

Product problems

  • Detachment Of Device Component
  • Component Falling

Patient

NA

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

AFTER COMPLETION OF AN EXAM, AND THE PT LEFT THE ROOM, THE OPERATOR ANGULATED THE TUBE FROM A HORIZONTAL TO A VERTICAL POSITION WHEN THE COLLIMATOR DETACHED AND FELL ONTO THE TABLE. THERE WAS NO INJURY REPORTED. THE CONCERN IS THAT A SERIOUS INJURY MAY RESULT IF THE FALLING COLLIMATOR WERE TO CONTACT THE PT OR OPERATOR.

Additional Manufacturer Narrative

A GE FIELD ENGINEER (FE) EVALUATED THE SYSTEM AND FOUND THAT ONE COLLIMATOR-TUBE INTERFACE SCREW WAS MISSING, ONE WAS LOOSE, AND ONE WAS SECURED. THE FE REATTACHED THE COLLIMATOR AND VERIFIED FUNCTIONALITY.