inforMED
MalfunctionJFP

ISE 9180

Received Mar 21, 2008 · Event occurred Feb 26, 2008

Report 1823260-2008-02656 · MDR key 1019675

Device

Generic name

Blood Gas Analyzer - Jfp

Manufacturer

Roche Diagnostics

Model number

ISE 9180

Catalog number

03157334001

Product problems

Patient

UNK

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

ACCOUNT EXPERIENCING GETTING ERRATIC PT RECOVERY FOR SODIUM STARTING 2007. NO PT RESULTS WERE PROVIDED. THE ONLY DATA PROVIDED WAS FROM A REPRODUCABILITY STUDY PERFORMED IN 2008, WHICH INCLUDED FOURTEEN SODIUM RESULTS FROM ONE PT SAMPLE. INITIAL RESULT GAVE 136 MMOL/L; SAMPLE REPEATED THIRTEEN TIMES GIVING 121, 135, 140, 137, 141, 138, 141, 132, 140, 111, 120, 137 & 132 MMOL/L. NO ADVERSE EVENTS REPORTED. IF ADDITIONAL INFO IS RECEIVED, APPROPRIATE NOTIFICATION WILL BE PROVIDED.