inforMED
MalfunctionHST

BUTTRESS/COMPRESSION NUT FOR 357.369

Received Jun 11, 2020

Report 2939274-2020-02784 · MDR key 10144803

Device

Generic name

Traction,apparatus,non-powered

Model number

357.371

Catalog number

357.371

Lot number

4987428

Product problems

  • Device-Device Incompatibility
  • Device-Device Incompatibility

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D4; D10; H4. H3, H6: THE DEVICE HAS BEEN RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. H11 CORRECTED DATA: G1: PHYSICAL MANUFACTURER. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. PART #: 357.371. SYNTHES LOT NUMBER: 4987428. MANUFACTURING SITE: (B)(4). RELEASE TO WAREHOUSE DATE: (B)(6) 2005. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. VISUAL INSPECTION: THE BUTTRESS/COMPRESSION NUT FOR 357.369 (PART # 357.371/ LOT # 4987428) WAS RECEIVED AT US CUSTOMER QUALITY (CQ). THE DISTAL PORTION OF THE BUTTRESS NUT WAS SEVERELY DEFORMED. THERE WERE SCRATCHES ALONG THE BODY OF THE NUT CONSISTENT WITH NORMAL WEAR. THE DEVICE COULD NOT BE REMOVED FROM THE MATED BLADE GUIDE SLEEVE FOR TROCHANTERIC FIXATION NAILS. SINCE THE DEVICE COULD NOT BE REMOVED, THE INTERNAL THREADS OF THE NUT COULD NOT BE INSPECTED. FUNCTIONAL TEST: FUNCTIONAL TESTING WAS PERFORMED WITH THE RETURNED DEVICES. THE BUTTRESS/COMPRESSION NUT WAS NOT ABLE TO BE DISASSEMBLED FROM THE MATED GUIDE SLEEVE. THE BUTTRESS NUT WAS NOT ABLE TO FURTHER LOOSEN/TIGHTEN. THE OVERALL COMPLAINT WAS CONFIRMED. CAN THE COMPLAINT BE REPLICATED WITH THE RETURNED DEVICE(S)? YES. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION COULD NOT BE CONFIRMED AS DEVICE DISASSEMBLY TO ACCESS INTERNAL COMPONENTS WAS NOT POSSIBLE. DOCUMENT/SPECIFICATION REVIEW: DRAWING(S) REVIEWED: CURRENT & MANUFACTURED REVISIONS CONCLUSION: THE OVERALL COMPLAINT WAS CONFIRMED FOR THE RECEIVED BUTTRESS/COMPRESSION NUT FOR 357.369 AS THE DEVICE WAS NOT ABLE TO DISASSEMBLE FROM THE GUIDE SLEEVE. ALTHOUGH NO DEFINITIVE ROOT-CAUSE CAN BE DETERMINED ITS POSSIBLE THE DEVICES HAVE GALLED TOGETHER OR ARE CROSS THREADED. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. NO DESIGN ISSUES WERE OBSERVED DURING THE DOCUMENT/SPECIFICATION REVIEW. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

ADDITIONAL NARRATIVE: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING AN UNKNOWN PROCEDURE, THE BUTTRESS COMPRESSION NUT FOR BLADE SLEEVE FOR TROCHANTERIC FIXATION NAILS GOT STUCK TOGETHER. THE BLADE GUIDE SLEEVE TROCHANTERIC FIXATION NAIL FAILED. THE BUTTRESS NUT AND GUIDE SLEEVE IMPACTED IN PATIENT. AN UNKNOWN MALLET WAS USED TO REMOVE THE DAMAGED SLEEVE AND BUTTRESS NUT. THERE WAS A SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN IMPACTION INST: HAMMER/MALLET: TRAUMA (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) BUTTRESS/COMPRESSION NUT FOR 357.369. THIS IS REPORT 1 OF 2 FOR (B)(4).