InjuryKPN
MALEM BEDWETTING ALARM
Received Jun 5, 2020 · Event occurred May 26, 2020
Report MW5094871 · MDR key 10128277
Device
Generic name
Alarm, Conditioned Response Enuresis
Manufacturer
Malem Medical Ltd.Model number
ULTIMATE BEDWETTING ALARMProduct problems
- Electrical /Electronic Property Problem
- Overheating of Device
- Defective Device
Patient
7 YR
- Injury
- Burn, Thermal
- Injury
- Burn, Thermal
Narrative
Description of Event or Problem
PATIENT IS A (B)(6) YEAR OLD MALE CHILD WHO WETS THE BED. PARENTS PURCHASED BEDWETTING ALARM FROM (B)(6) WEBSITE. THE ALARM HAD MALFUNCTION CAUSING INJURIES AND BURNS TO CHILD. CHILD WAS ADMITTED TO ER FOR BURNS (MINOR BUT SPREAD ON NECK). THE ALARM FAILED DUE TO DEFECT IN SHORT CIRCUIT ELECTRICAL PART. IT IS BELIEVED AS THE BATTERY HOLDER BENT FROM HEAT AND BATTERIES LEAKED ON CHILD'S CLOTHING AND ALL OVER THE ALARM. FDA SAFETY REPORT ID# (B)(4).