inforMED
MalfunctionGEC

BD SCRUB BRUSH E-Z SCRUB DRY STERILE

Received Jun 1, 2020 · Event occurred May 7, 2020

Report 1710034-2020-00340 · MDR key 10105599

Device

Generic name

Scrub Brush

Model number

371603

Catalog number

371603

Lot number

9184663

Product problems

  • Packaging Problem
  • Packaging Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED THAT 2 PACKAGES WERE DIFFICULT TO OPEN PRIOR TO USE AND WOULD RIP WITH A BD SCRUB BRUSH E-Z¿ SCRUB¿ DRY STERILE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 2 X PACKETS WERE DIFFICULT TO OPEN AND CAUSED RIPPING.

Description of Event or Problem

IT WAS REPORTED THAT 2 PACKAGES WERE DIFFICULT TO OPEN PRIOR TO USE AND WOULD RIP WITH A BD SCRUB BRUSH E-Z¿ SCRUB¿ DRY STERILE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 2 X PACKETS WERE DIFFICULT TO OPEN AND CAUSED RIPPING.

Additional Manufacturer Narrative

TWO SAMPLES RECEIVED. UNABLE TO OPEN EITHER SAMPLE, ALTHOUGH THE CUSTOMER HAD TRIED TO. COMPLAINT CONFIRMED. THE MOST LIKELY CAUSE WOULD BE FROM THE PACKAGING MACHINE STOPPING, AND RESIDUAL HEAT FROM THE SEALING DIE LOCALLY PRODUCING A SEAL THAT WAS TOO STRONG. ONE PRIOR COMPLAINT IN 2007 DHR REVIEWED. SEAL PEEL STRENGTH IS ONE OF THE CHARACTERISTICS MEASURE. ALL MEASUREMENTS WERE IN CONTROL LIMITS. H3 OTHER TEXT : SEE H.10.