COBAS® PCR MEDIA UNI SWAB SAMPLE KIT
Received May 28, 2020 · Event occurred May 3, 2020
Report 2243471-2020-00010 · MDR key 10097320
Device
Generic name
Specimen Collection Kit
Manufacturer
Roche Molecular Systems Inc.Model number
NACatalog number
07958030190
Lot number
F26931
Product problems
- Break
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THROUGH THE REPORTED CASE, IT WAS NOTED THAT THE COBAS PCR MEDIA UNI SWAB WAS NOT BROKEN PRIOR TO USE AND IT WAS LIKELY BROKEN AT THE SCORE-LINE BY THE PATIENT DURING SAMPLE COLLECTION DUE TO THEIR GAG REFLEX. INSTRUCTIONS EXIST IN THE COBAS PCR MEDIA UNI SWAB KIT INSTRUCTIONS FOR USE REGARDING PROPER SAMPLE COLLECTION. FACILITY NAME WAS TRUNCATED DUE TO CHARACTER LIMITATIONS. THE NAME OF THE FACILITY IS: (B)(4). (B)(4).
Description of Event or Problem
A CUSTOMER IN (B)(6) REPORTED A PATIENT SWALLOWED THE SHAFT OF THE COBAS PCR MEDIA UNI SWAB DURING COLLECTION OF A SPECIMEN FROM THE BILATERAL POSTERIOR PHARYNX. AS A CONSEQUENCE, THE PATIENT WAS BROUGHT INTO THE INTENSIVE CARE UNIT AND UNDERWENT TWO PROCEDURES (BRONCHOSCOPY AND ENDOSCOPY) TO LOCATE THE SWALLOWED SWAB SHAFT. THE SWAB SHAFT WAS LOCATED IN THE PATIENT'S UPPER THIRD OF THE ESOPHAGUS, AND WAS SUCCESSFULLY REMOVED. THE PATIENT IS NOW REPORTED AS DOING WELL.