Narrative

Additional Manufacturer Narrative

END USER'S WEIGHT NOT KNOWN SO ESTIMATION USED. TREND DATA REVIEWED AND NO ADVERSE TREND OBSERVED. BIOCOMPATIBILITY TESTING HAS BEEN PERFORMED IN ACCORDANCE WITH ISO 10993 AND MATERIALS WERE FOUND TO BE NON-SENSITIZERS. DHR REVIEW CONDUCTED AND RESULTS FOUND TO BE COMPLETE AND ACCURATE. SAMPLE REVIEW NOT POSSIBLE BECAUSE NO SAMPLE AVAILABLE. SINCE NYSTOP WAS PRESCRIBED FOR HER IRRITATED SKIN, THIS IS CONSIDERED AN MDR REPORTABLE EVENT. THE ROOT CAUSE OF THE EFFLUENT LEAKAGE CANNOT BE DETERMINED BUT THE PROCESS OF FINDING THE RIGHT BARRIER FIT MAY HAVE CONTRIBUTED TO THIS.

Description of Event or Problem

IT WAS REPORTED THAT AN END USER EXPERIENCED LEAKAGE UNDER HER HOLLISTER OSTOMY BARRIER WHICH LEAD TO SKIN IRRITATION. SHE WAS PRESCRIBED NYSTOP POWDER FOR HER IRRITATED SKIN AND IT HAS HELPED. HOLLISTER IS SENDING A SAMPLE OF A DIFFERENT BARRIER PRODUCT FOR HER TO TRY TO SEE IF IT WORKS BETTER.