MalfunctionNBC

ELECSYS PROBNP II IMMUNOASSAY

Received May 19, 2020 · Event occurred Apr 30, 2020

Report 1823260-2020-01257 · MDR key 10069827

Device

Generic name

Test,natriuretic Peptide

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

07027664190

Lot number

454331

Product problems

  • Low Test Results
  • Low Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE PATIENT SAMPLE COULD NOT BE PROVIDED FOR INVESTIGATION. THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS PROBNP II IMMUNOASSAY ON A COBAS 8000 E 801 MODULE. THE VALUES OBTAINED WITH THE PROBNP ASSAY DO NOT COMPARE TO A VALUE OBTAINED WITH A BNP ASSAY FROM AN UNKNOWN MANUFACTURER. THE SAMPLE IS SUSPECTED TO CONTAIN AN INTERFERENT TO A COMPONENT OF THE PROBNP ASSAY. THE VALUE FROM THE PROBNP ASSAY WAS REPORTED OUTSIDE OF THE LABORATORY. WHEN TESTED WITH THE ELECSYS PROBNP ASSAY ON THE E 801 ANALYZER, THE RESULT WAS < 5 PG/ML. THE SAMPLE WAS TESTED WITH A BNP ASSAY FROM AN UNKNOWN MANUFACTURER, RESULTING WITH A VALUE OF 565 PG/ML. THE SAMPLE WAS REPEATED WITH THE ELECSYS PROBNP ASSAY ON THE E 801 ANALYZER, RESULTING WITH A VALUE OF < 5 PG/ML. THE E 801 ANALYZER SERIAL NUMBER IS (B)(4).