SUTURE BUTTON 3.5MM, STERILE
Received Feb 29, 2008 · Event occurred Feb 5, 2008
Report 1220246-2008-00026 · MDR key 1005055
Device
Generic name
Implantable Device
Manufacturer
Arthrex, Inc.Model number
NACatalog number
AR-8920
Lot number
138043
Product problems
- Component(s), broken
Patient
UNK
- Surgical procedure, delayed
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR, THE SUTURE BROKE AS THE SURGEON WAS PASSING THE SUTURE THROUGH THE BUTTON. THE SURGERY WAS DELAYED APPROXIMATELY 45 MINUTES BUT NO ADD'L ADVERSE CONSEQUENCES WERE REPORTED WITH THE PT. THE SURGERY WAS COMPLETED WITH A DIFFERENT PRODUCT. THIS DEVICE IS USED FOR TREATMENT.
Additional Manufacturer Narrative
DEVICE EVALUATION REVEALED THAT ONLY THE BUTTON WAS RETURNED. A NEW SUTURE FROM STOCK WAS USED TO TEST AND PASSED THROUGH THE BUTTON WITHOUT PROBLEMS. THE SUTURE REMAINED INTACT. THE BUTTON HAD DEEP SCRATCHES AND GOUGES WHICH EVIDENCE CONTACT WITH A SHARP OBJECT. THE CAUSE OF THE COMPLAINANT'S EVENT, HOWEVER, COULD NOT BE DETERMINED FROM THE INFO AVAILABLE AND WITHOUT EVALUATION OF THE ACTUAL SUTURE INVOLVED.