PEPGEN P-15 FLOW
Received Feb 29, 2008 · Event occurred Jan 30, 2008
Report 1721411-2008-00007 · MDR key 1004596
Device
Product problems
- Failure to Osseointegrate
- Implant, removal of
Patient
60 YR
- Surgical procedure
Narrative
Description of Event or Problem
IT WAS REPORTED THAT PEPGEN P-15 FLOW WAS USED SIMULTRANEOUSLY WITH IMPLANT PLACEMENT IN THE RIGHT MAXILLA WITH BONE QUALITY TYPE II AND EXCELLENT ORAL HYGIENE. THE OSTEOTOMY WAS PREPARED VIA DRILLING AND BONE CONDENSING AND HEALING WAS TRANSGINGIVAL. THE IMPLANT DID NOT OSSEOINTEGRATE AND HAD SIGNS OF MOBILITY UPON EXPLANTATION DURING THE HEALING PERIOD, APPROX THREE WEEKS AFTER PLACEMENT. IT IS NOT CLEAR IF THE GRAFT WAS INTEGRATING WITH THE SURROUNDING VITAL BONE OR IF IT ALSO FAILED WITH THE IMPLANT.
Additional Manufacturer Narrative
FAILURE OF BONE GRAFTS TO OSSEOINTEGRATE IS AN INHERENT RISK OF THE PROCEDURE, ALTHOUGH UNCOMMON. OTHER VARIABLES, BEYOND PRODUCT NOT PERFORMING TO SPECS, TO CONSIDER IN A DIFFERENTIAL DIAGNOSIS WOULD BE PT HEALTH AND SOCIAL HISTORY (WHICH APPEAR TO BE UNREMARKABLE), SITE PREPARATION TRAUMA (HEAT OR PRESSURE) RESULTING IN NECROSIS OR SCARRING AND FIBROSIS, INSUFFICIENT PRIMARY STABILITY (SECONDARY TO BONE QUALITY AND/OR EXCESSIVE PREPARATION OF THE OSTEOTOMY AND QUANTITY OF MAXILLARY RIDGE BELOW THE MAXILLARY SINUS) OF THE IMPLANT RESULTING IN MICRO OR MACRO MOVEMENT AND SUBSEQUENT FIBROUS INTEGRATION, INDIVIDUAL GRAFTING TECHNIQUES, AND POST-OPERATIVE COMPLICATION (E.G. INFECTION). FROM THE INFO PROVIDED, IT IS NOT POSSIBLE TO DEFINITIVELY DETERMINE IF THE IMPLANT, GRAFT, TECHNIQUE, PT COMPLIANCE, POST-OPERATIVE COMPLICATION, ETC. WAS THE ETIOLOGY OF FAILURE. HOWEVER, BECAUSE A SECOND SURGERY WAS REQUIRED TO REMOVE AND/OR REPLACE THE IMPLANT AND GRAFT, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE IMPLANT LOSS WILL BE REPORTED VIA ASR, AS APPROPRIATE. THE DEVICE WAS NOT RETURNED FOR EVALUATION AND THE LOT NUMBER WAS NOT PROVIDED FOR RETAINED-PRODUCT TESTING AND/OR DHR REVIEW.