MalfunctionCAN
MEDICA CARBON DIOXIDE REGULATOR
Received May 4, 2020 · Event occurred Feb 9, 2020
Report 10020241 · MDR key 10020241
Device
Generic name
Regulator, Pressure, Gas Cylinder
Manufacturer
Western Medica / Scott Fetzer CompanyModel number
M1-940-PGCatalog number
M1-940-PG
Product problems
- Defective Component
Patient
0 DA
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
WHEN CHANGING CIRCUIT, ELEVATED CO2 NOTED, AND WOULD NOT COME DOWN UNTIL CO2 REGULATOR WAS CHANGED. MEDICA CARBON DIOXIDE REGULATOR. EQUIPMENT REMOVED FROM ECMO PUMP AND REPLACED WITH ANOTHER REGULATOR. INFANT HAD A SCHEDULED HEAD ULTRASOUND AFTER THE INCIDENT WHICH SHOWED NO CHANGE.