12,268 on file
- Carestation 620Jul 31, 2026Malfunction9710602-2026-01904
- Carestation 750Jul 31, 2026Malfunction9710602-2026-01905
- Carestation 750Jul 31, 2026Malfunction9710602-2026-01906
- Carestation 750cJul 31, 2026Malfunction9710602-2026-01908
- Carestation 750Jul 31, 2026Malfunction9710602-2026-01909
- Carestation 620Jul 31, 2026Malfunction9710602-2026-01910
- Carestation 650Jul 31, 2026Malfunction9710602-2026-01911
- Carestation 650Jul 31, 2026Malfunction9710602-2026-01912
- Carestation 620Jul 31, 2026Malfunction9710602-2026-01913
- Carestation 650Jul 31, 2026Malfunction9710602-2026-01914
Recalls
27 on file
- Open, ClassifiedJun 3, 2026
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
View FDA record ↗ - Open, ClassifiedDec 18, 2025
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
View FDA record ↗ - Open, ClassifiedDec 18, 2025
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
View FDA record ↗ - Open, ClassifiedDec 18, 2025
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
View FDA record ↗ - Open, ClassifiedJun 26, 2025
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
View FDA record ↗ - Open, ClassifiedJun 26, 2025
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
View FDA record ↗ - Open, ClassifiedApr 29, 2025
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
View FDA record ↗ - Open, ClassifiedApr 29, 2025
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
View FDA record ↗ - Open, ClassifiedApr 29, 2025
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
View FDA record ↗ - Open, ClassifiedApr 29, 2025
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
View FDA record ↗
Compliance history
Matched by FEI 3003513129.
FDA inspections
1 on file
- No Action Indicated (NAI)Jan 8, 2026
Compliance: Devices
Inspection citations
0 on file
No citations on file for this FEI number.
Compliance actions
0 on file
No warning letters or injunctions on file for this FEI number.
Import refusals
0 on file
No import refusals on file for this FEI number.