Wipro Ge Healthcare Private Ltd.
4, Kadugodi Industrial Area, Bangalore Karnataka, IN
Registered product codes
17 product codes
Recent MDR events
54 on file
- Lullaby WarmerJul 29, 2026Malfunction9617277-2026-00001
- GE MEDICAL NEONATAL PHOTOTHERAPY BILI BLANKETJun 29, 2026InjuryMW5189993
- MINI TELEMETRY SYSTEMMar 25, 2025Malfunction9617277-2025-00001
- MINI TELEMETRY SYSTEMMar 10, 2025MalfunctionMW5167445
- 250CX SERIESAug 24, 2023Death9617277-2023-00001
- GE COROMETRIC MINI TELEMETRYDec 30, 2022MalfunctionMW5114111
- IVENT 201Nov 29, 2022Malfunction3007734888-2022-00001
- IVENT 201Nov 29, 2022Malfunction3007734888-2022-00002
- RADIANT WARMEROct 13, 2022MalfunctionMW5112644
- GE CARESTATION OMNIBED INCUBATORJan 28, 2022InjuryMW5107073
Recalls
10 on file
- Open, ClassifiedOct 21, 2025
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
View FDA record ↗ - Open, ClassifiedOct 21, 2025
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
View FDA record ↗ - Open, ClassifiedSep 20, 2024
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
View FDA record ↗ - Open, ClassifiedSep 20, 2024
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
View FDA record ↗ - Open, ClassifiedFeb 9, 2024
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
View FDA record ↗ - Open, ClassifiedFeb 9, 2024
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
View FDA record ↗ - Open, ClassifiedFeb 9, 2024
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
View FDA record ↗ - Open, ClassifiedFeb 9, 2024
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
View FDA record ↗ - Open, ClassifiedFeb 9, 2024
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
View FDA record ↗ - Open, ClassifiedFeb 9, 2024
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
View FDA record ↗
Compliance history
Matched by FEI 3003190524.
FDA inspections
0 on file
No inspections on file for this FEI number.
Inspection citations
0 on file
No citations on file for this FEI number.
Compliance actions
0 on file
No warning letters or injunctions on file for this FEI number.
Import refusals
0 on file
No import refusals on file for this FEI number.